Omron Bp 485 User's Manual

KNOW YOUR UNIT Main Unit: Display Battery compartment START/STOP button Memory (M) button Air jack Components: Air tube INSTRUCTION MANUAL Automatic Blood Pressure Monitor Model BP710 INTRODUCTION Thank you for purchasing the OMRON® BP710 Automatic Blood Pressure Monitor. Fill in for future reference. Contents ® Thank you for purchasing the OMRON BP786 Blood Pressure Monitor. Your new blood pressure monitor uses the oscillometric method of blood pressure measurement. This means the monitor detects your blood movement through your brachial artery and converts the movements into a digital reading.

Wrist Blood

Pressure Monitor

Model RS2

Instruction Manual

IM-HEM-6121-E-02-01/2013 5338844-7B

Introduction

Thank you for purchasing the OMRON RS2 Wrist Blood Pressure Monitor.

The OMRON RS2 is a compact and easy to use blood pressure monitor, operating on the oscillometric principle. It measures your blood pressure and pulse rate simply and quickly. For comfortable controlled inflation without the need of pressure pre-setting or re-inflation the device uses its advanced “IntelliSense” technology.

Intended Use

This product is designed to measure the blood pressure and pulse rate of people within the range of the designated wrist cuff, following the instructions in this instruction manual.

It is mainly designed for general household use. Please read the Important Safety Information in this instruction manual before using the unit.

Please read this instruction manual thoroughly before using the unit.

Please keep for future reference.

For specific information about your own blood pressure, CONSULT YOUR DOCTOR.

Important Safety Information

Consult your doctor prior to using in pregnancy or if diagnosed with arrhythmia or arteriosclerosis.

Please read this section carefully before using the unit.

Warning:

• Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.

(General Usage)

Always consult your doctor. Self-diagnosis of measurement results and self-treatment are dangerous.

People with severe blood flow problems, or blood disorders, should consult a doctor before using the unit, as cuff inflation can cause internal bleeding.

(Battery Usage)

If battery fluid should get in your eyes, immediately rinse with plenty of clean water. Consult a doctor immediately.

Caution:

• Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury to the user or patient or damage to the equipment or other property.

(General Usage)

Do not leave the unit unattended with infants or persons who cannot express their consent.

Do not use the unit for any purpose other than measuring blood pressure.

Do not use a mobile phone or other devices that emit electromagnetic fields, near the unit. This may result in incorrect operation of the unit.

Do not disassemble the unit or wrist cuff.

Do not operate the unit in a moving vehicle (car, airplane).

(Battery Usage)

If battery fluid should get on your skin or clothing, immediately rinse with plenty of clean water.

Use only two “AAA” alkaline (LR03) batteries with this unit.

Do not insert the batteries with their polarities incorrectly aligned.

Replace old batteries with new ones immediately. Replace both batteries at the same time.

Remove the batteries if the unit will not be used for three months or more.

Do not use new and used batteries together.

General Precautions

Do not apply strong shocks and vibrations to or drop the unit.

Do not take measurements after bathing, drinking alcohol or caffeine, smoking, exercising or eating.

Do not inflate the wrist cuff when it is not wrapped around your wrist.

Read and follow the “Important information regarding Electro Magnetic Compatibility (EMC)” in the Technical Data Section.

Read and follow the “Correct Disposal of This Product” in the Technical Data Section when disposing of the device and any used accessories or optional parts.

1. Overview

Main unit:

A

D

B

C

E

A. Wrist cuff

D. Memory ( ) button

B. Battery compartment

E. START/STOP button

C. Display

Display:

F

J

K

L

G

H

M

N

I

F. Systolic blood pressure

J. Memory symbol

G. Diastolic blood pressure

K. Irregular heartbeat symbol

H. Heartbeat symbol

L. Cuff wrapping guide

1. Flashes during measurement

M. Deflation symbol

2. If flashing after measurement

N. Battery low symbol

completed or when viewing

results stored in the memory,

* Note: If your systolic or diastolic

indicates blood pressure out of

recommended range*.

pressure is outside the standard

range (above 135/85 mmHg) the

I. Pulse display and Memory

Heartbeat symbol ( ) will blink.

number

Please refer to Section 3.3.

2.Preparation

2.1Installing/Replacing the Batteries

1. Remove the battery cover.

2.Insert two 1.5V “AAA” alkaline (LR03) batteries as indicated in the battery

compartment and then replace the battery cover.

Notes:

If the battery low symbol () appears on the display, turn off the unit then replace both batteries at the same time.

The measurement values continue to be stored in memory even after the batteries are replaced.

Disposal of used batteries should be carried out in accordance with the national regulations for the disposal of batteries.

3.Using the Unit

3.1Applying the Wrist Cuff

Do not apply over clothing.

1. Place the wrist cuff over your wrist.

Your palm should face upward.

2. Wrap the wrist cuff around your wrist.

Wrap the wrist cuff securely around the wrist for taking accurate measurements.

Make sure that the wrist cuff does not cover the protruding part of the wrist bone (ulna) on the outside of the wrist.

Note: You can take a measurement on either your left or right wrist.

3.2 How to Sit Correctly

To take a measurement, you need to be relaxed and comfortably seated, under comfortable room temperature. No bathing, drinking alcohol or caffeine, smoking, exercising or eating 30 minutes before taking a measurement.

• Sit on a chair with your feet flat on the floor.

• Sit upright with your back straight.

The cuff should be at the same level as your heart.

Relax your wrist and hand. Do not bend your wrist back, clench your fist, or bend your wrist forward.

3.3 Taking a Reading

Notes:

To cancel a measurement, press the START/STOP button at any time during measurement.

Remain still while taking a measurement.

1.Press the START/STOP button.

User

All the symbols appear on the display.

The wrist cuff will start to inflate automatically.

Cuff Wrapping Guide

START INFLATE DEFLATE COMPLETE

Cuff Wrapping Guide

The Cuff Wrapping Guide is a unique feature that indicates if the cuff is not wrapped tightly enough around the wrist. Even when the is displayed, a blood pressure reading will be taken.

Note: This reading is NOT reliable due to the incorrect wrapping of the cuff. Please wrap the cuff again, taking care to wrap it correctly and take the measurement

again. When the is displayed, the cuff is correctly wrapped tightly enough on the wrist and the reading is accurate and reliable.

2.Undo the wrist cuff and remove the unit.

3.Press the START/STOP button to turn off the monitor.

The monitor automatically stores the measurement in its memory.

It will automatically turn off after two minutes.

Important:

• If your systolic or diastolic pressure is outside the standard range, the heartbeat symbol will blink when the measurement result is displayed.

Manual

Recent research suggests that the following values can be used as a guide to high blood pressure for measurements taken at home.

Systolic Blood Pressure

Above 135 mmHg

Diastolic Blood Pressure

Above 85 mmHg

This criteria is for home blood pressure measurement.

• Your blood pressure monitor includes an

irregular heartbeat feature. Irregular heartbeats can influence the results of the

measurement. The irregular heartbeat algorithm automatically determines if the measurement is usable or needs to be repeated. If the measurement results are affected by irregular heartbeats but the result is valid, the result is shown together with the

irregular heartbeat symbol ().

If the irregular heartbeats cause the measurement to be invalid, no result is shown. If the irregular heartbeat symbol () is shown after you have taken a measurement, repeat the measurement. If the irregular heartbeat symbol () is shown frequently, please make your doctor aware of it.

Notes:

Wait 2 - 3 minutes before taking another blood pressure measurement. Waiting between readings allows the arteries to return to the condition prior to taking the blood pressure measurement.

The blood pressure can differ between the right arm and the left arm, and therefore also the measured blood pressure values can be different. OMRON recommends to always use the same arm for measurement. If the values between the two arms differ substantially, please check with your physician which arm to use for your measurement.

Warning:

Self-diagnosis of measured results and treatment are dangerous. Please follow the instructions of your doctor.

3.4 Using the Memory Function

The unit automatically stores up to 30 sets of measurement values (blood pressure and pulse rate). If the memory is full, the monitor will delete the oldest readings.

To View the Readings Stored in Memory

1. Press the button.

The Memory number appears for a second before the pulse rate is displayed. The newest set is numbered “1”.

Note: If there are no measurements results stored in the memory, the screen to the right is displayed.

2.Press the button repeatedly to view the readings stored in memory.

To Delete All the Values Stored in Memory

When the memory symbol () appears, first press the button. Then while holding it down, press the START/STOP button simultaneously for about 2 - 3 seconds.

First Second

Note: You cannot partially delete the stored readings.

4.Troubleshooting and Maintenance

4.1The Icons and Error Messages

Error Display

Cause

Remedy

Remove the wrist cuff. Wait

2 - 3 minutes and then take

Irregular heartbeats

another measurement.

are detected.

Repeat the steps in section

3.3. If this error continues to

appear, contact your doctor.

The batteries are

You should replace them with

Blink

new ones ahead of time.

low.

Refer to section 2.1.

The batteries are

You should replace them with

Lit

new ones at once.

exhausted.

Refer to section 2.1.

Wrist cuff not

Apply the wrist cuff correctly.

applied correctly.

Refer to section 3.1.

Repeat Measurement while

Movement during

remaining still and refraining

from talking during the

measurement.

measurement.

Refer to section 3.3.

Apply the wrist cuff correctly

Wrist cuff not

and repeat measurement

applied correctly, or

while remaining still and

movement during

refraining from talking during

measurement.

the measurement.

Refer to section 3.1 and 3.3.

Device error.

Contact your OMRON retail

outlet or distributor.

Note: The irregular heartbeat symbol ( ) may also be displayed with error messages.

4.2 Troubleshooting

Problem

Cause

Remedy

The wrist cuff is not

Measure while in the

at heart level.

correct posture.

Refer to section 3.2.

The cuff is not

Wrap the cuff correctly.

wrapped snugly

The reading is

Refer to section 3.1.

around the wrist.

extremely low

The arms and

Relax and try taking the

(or high).

shoulders are

measurement again.

tense.

Refer to section 3.3.

Movement or

Remain still and do not

talking during

talk during Measurement.

measurement.

Refer to section 3.3.

Wrist cuff pressure

Air is leaking from

Consult your OMRON

does not rise.

the wrist cuff.

retail outlet or distributor.

Apply the cuff correctly so

Wrist cuff deflates

The wrist cuff is

that it is firmly wrapped

too soon.

loose.

around the wrist. Refer to

section 3.1.

Blood pressure readings

constantly vary with time

The blood pressure is different

of day and how relaxed

each time. The reading is

you are. Take several

extremely low (or high).

deep breaths and try to

remain relaxed before

taking a measurement.

The unit loses

Replace the batteries with

power during

The batteries are

measurement.

new ones.

exhausted.

Refer to section 2.1.

Nothing happens

when you press

The batteries have

Insert the batteries with

the buttons.

been inserted

the correct (+/-) polarity.

incorrectly.

Refer to section 2.1.

• Press the START/STOP button and repeat

measurement.

Other problems.

• If the problem continues, try replacing the

batteries with new ones.

If this still does not solve the problem, contact

your OMRON retail outlet or distributor.

4.3 Maintenance

To protect your unit from damage, please observe the following:

•Do not subject the main unit and the cuff to extreme temperatures, humidity, moisture or direct sunlight.

•Do not disassemble the unit.

•Do not subject the unit to strong shocks or vibrations (for example, dropping the unit on the floor).

•Do not use volatile liquids to clean the main unit. •Do not wash the cuff or immerse it in water. •Do not use petrol, thinners or similar solvents to clean the cuff.

•Do not carry out repairs of any kind yourself. If a defect occurs, consult your OMRON retail outlet or distributor as mentioned on the packaging.

•The unit should be cleaned with a soft, dry cloth. •Use a soft, moistened cloth and soap to clean the cuff.

•Keep the unit in its storage case when not in use.

•Fold the cuff into the storage case.

Do not store the unit in the following situations:

•If the unit is wet.

•Locations exposed to extreme temperatures, humidity, direct sunlight, dust or corrosive vapours.

•Locations exposed to vibrations, shocks or where it will be at an angle.

Calibration and Service

•The accuracy of this blood pressure monitor has been carefully tested and is designed for a long service life.

•It is generally recommended to have the unit inspected every two years to ensure correct functioning and accuracy. Please consult your authorised OMRON dealer or the OMRON Customer Service at the address given on the packaging or attached literature.

5. Technical Data

Product Description

Wrist Blood Pressure Monitor

Model

OMRON RS2 (HEM-6121-E)

Display

LCD Digital Display

Measurement Method

Oscillometric method

Measurement Range

Pressure: 0 to 299 mmHg

Pulse: 40 to 180/min.

Accuracy

Pressure: ±3 mmHg

Pulse: ±5% of display reading

Inflation

Automatic inflation by pump

Deflation

Automatic rapid deflation

Memory

30 Measurements

Power Source

2 “AAA” alkaline (LR03) batteries 1.5V

Battery Life

Approx. 300 measurements with new

alkaline batteries at a room temperature of

23°C

Applied Part

= Type B

Protection Against

Internally powered ME equipment

Electric Shock

Operating temperature/

+10 to +40°C / Maximum: 30 to 85% RH

Humidity

Storage temperature/

-20 to +60°C / Maximum: 10 to 95% RH /

Humidity

700 to 1060hPa

Console Weight

Approximately 101g without batteries

Outer Dimensions

Approximately 78 (w) mm x 60 (h) mm x

21 (d) mm (without the wrist cuff)

Measurable

Approximately 13.5 to 21.5 cm

circumference

Cuff Material

Nylon and polyester

Package Content

Main unit, storage case, battery set,

instruction manual, guarantee card

Note: Subject to technical modification without prior notice.

This device fulfils the provisions of EC directive 93/42/EEC (Medical Device Directive).

This blood pressure monitor is designed according to the European Standard EN1060, Non-invasive sphygmomanometers Part 1: General Requirements and Part 3: Supplementary requirements for electromechanical blood pressure measuring systems.

This OMRON product is produced under the strict quality system of OMRON HEALTHCARE Co. Ltd., Japan. The Core component for OMRON blood pressure monitors, which is the Pressure Sensor, is produced in Japan.

Important information regarding Electro Magnetic Compatibility (EMC)

With the increased number of electronic devices such as PC’s and mobile (cellular) telephones, medical devices in use may be susceptible to electromagnetic interference from other devices. Electromagnetic interference may result in incorrect operation of the medical device and create a potentially unsafe situation. Medical devices should also not interfere with other devices.

In order to regulate the requirements for EMC (Electro Magnetic Compatibility) with the aim to prevent unsafe product situations, the EN60601-1-2:2007 standard has been implemented. This standard defines the levels of immunity to electromagnetic interferences as well as maximum levels of electromagnetic emissions for medical devices.

This medical device manufactured by OMRON HEALTHCARE conforms to this EN60601-1-2:2007 standard for both immunity and emissions.

Nevertheless, special precautions need to be observed:

Do not use mobile (cellular) telephones and other devices, which generate strong electrical or electromagnetic fields, near the medical device. This may result in incorrect operation of the unit and create a potentially unsafe situation. Recommendation is to keep a minimum distance of 7 m. Verify correct operation of the device in case the distance is shorter.

Further documentation in accordance with EN60601-1-2:2007 is available at OMRON HEALTHCARE EUROPE at the address mentioned in this instruction manual.

Documentation is also available at www.omron-healthcare.com.

Correct Disposal of This Product

(Waste Electrical & Electronic Equipment)

This marking shown on the product or its

literature, indicates that it should not be disposed of, with other household wastes at the

end of its working life.

To prevent possible harm to the environment or human health from uncontrolled waste disposal, please separate this product from other types of wastes and recycle it responsibly to promote the sustainable reuse of material resources.

Omron Bp 485 User's Manual Diagram

Household users should contact either the retailer where they purchased this product, or their local government office, for details of where and how they can return this item for environmentally safe recycling.

Omron 10 User Manual

Business users should contact their supplier and check the terms and conditions of the purchase contract. This product should not be mixed with other commercial wastes for disposal.

Manufacturer

OMRON HEALTHCARE Co., Ltd.

53, Kunotsubo, Terado-cho, Muko, Kyoto,

617-0002 JAPAN

Omron Bp 485 User's Manual Instructions

EU-representativeOMRON HEALTHCARE EUROPE B.V.

Scorpius 33, 2132 LR Hoofddorp THE NETHERLANDS www.omron-healthcare.com

Production facility OMRON (DALIAN) CO., LTD.

Omron Bp Instructions

Dalian, CHINA

Subsidiary

OMRON HEALTHCARE UK LTD.

Opal Drive, Fox Milne, Milton Keynes, MK15 0DG

U.K.

OMRON MEDIZINTECHNIK

HANDELSGESELLSCHAFT mbH

John-Deere-Str. 81a, 68163 Mannheim, GERMANY

www.omron-medizintechnik.de