Manual General Us Voluson - Free ebook download as PDF File (.pdf), Text File (.txt) or read book online for free. Voluson 730 Pro V Brochure. Fuji FCR 5000 Service Manual. Key on the control panel, or the X symbol on the Help window to exit the electronic user manual. Voluson 730Expert - Basic User Manual.


voluson 730

July 18, 2012 03:09

i have voluson 730 , bt02 , this system do not start up just writ on touch screen BOOTING , and on big screen try to start up and return from beginning i hope some body help me to start it to work

Log inor Register

to rate and post a comment

I have the service manual in PDF SEND ME YOUR EMAIL

Log inor Register

to rate and post a comment

Voluson 730 pro service manual

Kindly send me this manual at khurramnisarsiddiqi@yahoo.com

Log inor Register

to rate and post a comment

this problem is for bep or main board.please check the sbc or main boRd.some times battery bord is defective.after it check the power cabel.

Log inor Register

to rate and post a comment

Hi Friend, If you can send a copy of manual i appreciatte. Thanks... mauriciocidmx@hotmail.com

Log inor Register

to rate and post a comment

Rafael, I am trying to service my bios battery and am getting no help from the tech support, since they want to get paid to do it for me. Can you send me your service manual PDF so I can quit messing with them. rtulley@hotmail.com. thank you in advance. Pam Tulley

Log inor Register

to rate and post a comment

probably the BIOS/CMOS Battery ran out, you need to reaplce and load the right settigns on the BIOS, I have seen these kind of problems before.

Log inor Register

to rate and post a comment

Some how the SBC board in our Voluson 730 (4D) does not have any battery. I looked at the youtube video as well the GE Service manual and the SBC Board shown in them are different than I have. Our machine was bought in 2001.
I would like to attach a photo of the SBC Board to this email but cannot find any option to do so.
Any help would be appreciated from any of the service engineers.

Log inor Register

to rate and post a comment

I would suggest to reload software. Just remember the BT02 systems use a boot CD to start the load procedure.

Log inor Register

to rate and post a comment

hello
thank you for your help , i have CD for software , but how i can reload it ?

Log inor Register

to rate and post a comment

First do the BIOS check thats really well explain there. If that doesn't work then the SW load will be the next step. Let me see if I have the loading procedure.

Log inor Register

to rate and post a comment

https://www.youtube.com/watch?v=VvmLiN1zj3Y&list=PL5196474BA82113E3&index=1
here is video how to reload software, when it boot you have to enter the bios pass its Rudi, you have to select the system to boot with an optical device.

Log inor Register

to rate and post a comment

I have encountered a similar issue and Mr. Sanchez is likely correct. It should be easy to check however. As the system begins to boot you need to enter the BIOS settings. The system does not tell you what key to press but it is either 'F1', 'F2', or 'Del'. The best thing to do is to press each one in sequence repeatedly until the BIOS screen is displayed. Once this comes up, choose the option to 'load defaults'. Do this and then choose the option to 'save settings and exit'. The system should then reboot normally.
If this works, then the problem is definitely the BIOS battery. It is a CR2032 coin battery that is located on the SBC (Single Board Computer) inside the system cabinet. Opening the system cabinet will require opening the back of the system, disconnecting all of the cables attached to it and pulling the entire box out. Voluson did not make it easy.
Good luck with this.

Log inor Register

to rate and post a comment

really thank you , i will do what you inform me.

Log inor Register

to rate and post a comment

i do what you said and press many times F1 , F2 , DELL but system do not go to bios , but if we press ctrl&a , system go to SCSI .
when system start up write on screen SCSI BIOS NOT INSTALLED

Log inor Register

to rate and post a comment

The 'SCSI BIOS NOT INSTALLED' message is of no matter. Are you pressing the F1, F2, Del as soon as you turn the system on? Each button must be pressed separately.

Log inor Register

to rate and post a comment

as sson as the system starts press the Key DEL, that will take you to the BIOS, yit may ask you for a password, Rudi

Log inor Register

to rate and post a comment

I have a voluson 730 expert BT01 the system does boot ok, but touch panel not working, only work keyboard and mouse
What can I do?
Thanks

Log inor Register

to rate and post a comment

Dear
you must chek the main pord of the pc or sbc bord.
it is error for main bord or sbc bord

Log inor Register

to rate and post a comment

Yeah. You can try to take SBC out and clean it well. Reseat it and try again. Sometimes those boards get bad connections.

Log inor Register

to rate and post a comment

hi
i enter to BIOS but i do not know what i must do , any way this system try to start but do not go on and blue screen appear written (beginning dump of physical memory
physical memory a )
so are there problem in memory?

Log inor Register

to rate and post a comment

When on BIOS, you have to load optimal configuration and save and exit. After that the system might boot without problem.

Log inor Register

to rate and post a comment

When you are in the BIOS, there will be a selection to load the default setup. Choose that. There will then be a selection to save the settings and exit. Choose that. It will probably ask for you to confirm this (Y/N). Type 'Y'. The system should then reboot and hopefully come up normally. If it does, then you have only fixed the problem temporarily. The BIOS battery on the SBC will need to be replaced.

Log inor Register

to rate and post a comment

re: re: re: re: re: voluson 730

July 24, 2012 12:41

thank you for your help, i did that but still same problem the system do not go on start to go but return start again

re: re: re: re: re: voluson 730

July 24, 2012 12:41

hello
i did but the problem still same and system do not go on

re: re: re: re: re: re: voluson 730

July 24, 2012 05:31

Voluson 730 Pro User Manual Pdf

I think that it is possible then that either the operating system is corrupt of the system hard drive has failed.

re: re: re: re: re: re: re: voluson 730

July 25, 2012 10:02

OH, CAN WE CHANGE IT EASY AND HOW WE CAN FIND IT?

re: re: re: re: re: re: re: re: voluson 730

July 25, 2012 12:50

If reloading software is possible, I would try that first. Replacing anything on this system is a challenge. Everything is contained within the large stainless steel box within the base of the unit. All of the wires and cables must be disconnected from this box (mark which ones plug into where) and it must be removed from the unit before it can be opened. Make sure that all of the probes have been removed. All of the system components including the hard drive are contained within.

re: re: re: re: re: re: voluson 730

July 25, 2012 10:14

Reload SW would be your best option now.

re: re: re: re: re: re: re: voluson 730

Voluson I Manual

July 25, 2012 12:50

i have two CD for software but how we can reload it i put cd in cd reader but no think happen

re: re: re: re: re: re: re: re: voluson 730

July 25, 2012 02:56

You have to change the BIOS setting to boot with the CD.

re: re: re: re: re: re: re: voluson 730

July 25, 2012 12:50

One important thing you need to have when reloading SW on the system, is to have all your options listed with their keys.

Ge Voluson E6 User Manual

Log inor Register

to rate and post a comment

re: re: re: re: re: re: re: re: voluson 730

March 01, 2013 02:17

where is the key

re: re: re: re: re: re: re: re: re: voluson 730

March 01, 2013 05:24

You have to go to settings, and then system. You should find options listed there. Print them or copy them.

re: re: re: re: re: re: re: re: re: re: voluson 730

March 22, 2013 06:00

how I enter to setting and then system
I have voluson 730 expert BT01
I have software cd but no boot cd

re: re: re: re: re: re: re: re: re: re: re: voluson 730

March 23, 2013 06:44

I have voluson 730 BT01
the system do not go on start to go but return start again
The BIOS battery on the SBC is new

frend.i check your text.do you want work your system or no.i see some body say some ting that all is not curect.check step by step.1.open the back door.2.check the 6 lamp of voltag.if 6 of them is ok.if not ok and 1 or some of them be off check the pwer supplay.if is ok open the Bep and check the voltag by voltag cheker for the moder bord.if all voltag is not ok the cpp bord is de fectiv.if you have all reng of voltag and evry thing is ok the main bord i100% is defective.cheng it.
yousefi

Log inor Register

to rate and post a comment

I have a voluson 730 expert BT01, the system does boot ok but touch panel not working
What can I do?
Thanks

Log inor Register

Voluson 730 Pro Service Manual

to rate and post a comment

GE Voluson 730 Expert - 'OCP Connection Failed' Error shows up after five attempts of 'Connecting to Console'
Can someone help finding solution on this? Thanks

Log inor Register

to rate and post a comment

you have connected something wrong, the control panel can't connect to the GET box

Log inor Register

to rate and post a comment

One moment please...
  • Parts
  • Videos
  • News
DESCRIPTION
The Voluson 730 PRO demonstrates GE Healthcare0019s commitment to expanding the clinical impact of 3D/4D technology through education and ergonomics.

Obstetrics

The Voluson 730 PRO provides a full suite of clinical tools including multiplanar analysis, Volume Calculation (VOCAL), Compound Resolution Imaging (CRI) and image optimization techniques. These tools aid in the analysis of a variety of fetal anomalies such as cleft lip and palate, spina bifida, club foot and trisomy 21.

Gynecology

Automatic volume acquisition and multiplanar analysis facilitate saline-infused sonography while multiplanar & rendering capabilities enable evaluation of the shape of the uterine cavity.

Manual

Breast

Multiplanar views and RealTime 4D biopsy increase confidence of breast biopsy through visualization of needle placement.

FORUMSView All (33)
Ask a New Question

0

Replies
-toridi
3 months ago
3 months agoCPP81-82.P3 Power Supply Board
CPP81-82.P3 Power Supply Board, how check condition power supply in outside , ? i need datasheet this power supplythks beforeToriditoridi_fondaco@yahoo.comReply
-Babygram
3 months ago
3 months agoSaving images to USB flashdrive on Voluson 730 Pro
Hi there! Is there a way to save the images on a USB flashdrive instead of a CD? Reply

1

Reply
-Vittals Medicare
a year ago
a year agostandby switch
Standby switch connection formatReply
DOCUMENTS / MANUALSView All

FEATURES

  • Full DICOM functionality
  • 3D/4D functionality in a PACS environment
  • 80GB hard drive with DVD/CD archiving capabilities

SPECIFICATIONS

Cart Based1
Clinical ApplicationsColorectal
Color-flow Mapping1
Doppler Modes3-D, Continuous wave (CW), Pulsed wave (PW), HPRF
Height55.9 in
Image Display15 in Color Flat Panel
Imaging Capabilities3D/4D
Length 39.4 in
Weight300 lbs
Width26.7 in